Development of a lung nodule volumetry tool for quality assurance assessment of low-dose lung cancer screening CT protocols.
Low-dose CT (LDCT) is the preferred modality for lung nodule detection and follow-up, yet standardized quality-control criteria are lacking. Volumetric accuracy is clinically important but strongly dependent on segmentation and reconstruction settings, particularly for small, low-contrast subsolid nodules, and frameworks for optimizing LDCT protocols based on volumetric performance remain limited.
To develop a lung nodule segmentation and volumetry tool that evaluates volumetric accuracy across CT reconstruction settings and to demonstrate how accuracy and repeatability metrics can support protocol comparison as a quality-assurance assessment.
A custom semi-automatic tool was developed in Python and evaluated using the QRM Thorax phantom on a GE Revolution Apex CT system. Fifteen artificial nodules were analyzed, combining five diameters (3-12 mm) and three attenuation levels (100, -630, -800 HU). Reconstruction parameters were systematically varied, including kernel (STANDARD, LUNG), slice thickness (0.625-2.5 mm), spacing (0.625, 1.25 mm), algorithm (DLIR-H and ASiR-V at 80%), and field of view (18, 36 cm). Performance was assessed using mean relative volume error (target within± 20%), repeatability coefficient (RC), and size-specific RMSE; the best-performing configuration per nodule type served as reference. Reconstruction effects were evaluated using factorial ANOVA, Levene's test, Welch's t-test, and bootstrap analysis with false-discovery-rate correction.
Overall mean relative volume errors ranged from -7.1% to 6.7% for solid nodules and -14.7% to 5.8% for type 1 subsolid nodules≥ 5 mm; type 2 subsolid nodules showed larger deviations (-27.1% to -12.1%). Nodule size and reconstruction kernel were the dominant contributors to mean volume error, while slice thickness and spacing mainly influenced variability. Significant differences from the reference were found for four solid-nodule settings, none for type 1, and one for type 2 subsolid settings, with degradation occurring mainly at increased slice thickness or reduced field of view. The 3 mm subsolid nodules showed consistently high variability across all settings.
The tool enabled quantitative comparison of CT reconstruction settings using volumetric accuracy and repeatability metrics. For clinically relevant sizes (≥ 5 mm), volumetric performance remained within the predefined technical-suitability range across most settings, supporting use of the tool for protocol evaluation and future dose-optimization studies.
To develop a lung nodule segmentation and volumetry tool that evaluates volumetric accuracy across CT reconstruction settings and to demonstrate how accuracy and repeatability metrics can support protocol comparison as a quality-assurance assessment.
A custom semi-automatic tool was developed in Python and evaluated using the QRM Thorax phantom on a GE Revolution Apex CT system. Fifteen artificial nodules were analyzed, combining five diameters (3-12 mm) and three attenuation levels (100, -630, -800 HU). Reconstruction parameters were systematically varied, including kernel (STANDARD, LUNG), slice thickness (0.625-2.5 mm), spacing (0.625, 1.25 mm), algorithm (DLIR-H and ASiR-V at 80%), and field of view (18, 36 cm). Performance was assessed using mean relative volume error (target within
Overall mean relative volume errors ranged from -7.1% to 6.7% for solid nodules and -14.7% to 5.8% for type 1 subsolid nodules
The tool enabled quantitative comparison of CT reconstruction settings using volumetric accuracy and repeatability metrics. For clinically relevant sizes (