Use of the MemoSorb Bioresorbable PFO Device in an 11-Year-Old Diver with Phospholamban Mutation and Familial Hypertrophic Cardiomyopathy.

We report biodegradable device implantation for primary prevention in an asymptomatic 11-year-old boy with a patent foramen ovale, normal biventricular function and phospholamban mutation, conferring an elevated future risk of hypertrophic cardiomyopathy. He was a keen diver with ambitions to be a marine biologist. Following multidisciplinary discussion, closure of the defect was agreed using a bioresorbable device to optimize future transseptal intervention for potential cardiomyopathy. A 34 mm MemoSorb® PFO Occluder (Lepu medical Technology, Beijing, China) was positioned, complicated by complete atrioventricular block, regressing to first degree after removing the device. A 28 mm device was then successfully implanted. Intermittent 2nd degree atrioventricular block occurred post-procedurally but was successfully treated with steroids, with normal Holter ECG a few weeks later. Bubble test echocardiography was negative after 6 months. The patient was allowed to dive again. The MemoSorb ® biodegradable device can be implanted in selected cases. More studies are needed to confirm safety and efficacy.